MechanicalVernier & Digital Calipers
Vernier, dial and digital calipers up to 600 mm, plus depth and gear tooth verniers.
What we calibrate, with the exact ranges and calibration capability from our scope of accreditation.
Instruments
Filter by discipline or search for your instrument.
MechanicalVernier, dial and digital calipers up to 600 mm, plus depth and gear tooth verniers.
MechanicalExternal, internal, depth, three-point and V-anvil micrometers.
MechanicalPlunger and lever dial gauges, bore gauges, thickness gauges and height gauges.
MechanicalAnalog and digital pressure gauges and transducers up to 700 bar (on-site).
MechanicalElectronic balances from 210 g to 100 kg, as per OIML R 76-1.
MechanicalContact and non-contact tachometers up to 38,000 rpm, sound level meters.
Electro-TechnicalDigital multimeters and clamp meters: AC/DC voltage, current (up to 1000 A) and resistance.
Electro-TechnicalPower supplies, multifunction calibrators, resistance boxes, panel meters and insulation testers.
ThermalRTDs, thermocouples, controllers, data loggers and liquid-in-glass thermometers.
ThermalTemperature and humidity indicators, controllers and chambers (20–90 %RH).
ThermalInfrared guns and pyrometers from 30 °C to 500 °C.
ThermalSingle and multi-position temperature mapping from −80 °C to 1200 °C.
MechanicalClick-type, dial and digital torque wrenches, screwdrivers and torque testers.
MechanicalPush-pull force gauges, load cells and force indicators.
MechanicalCompression and universal testing machines, calibrated at your site.
MechanicalChecking fixtures, jigs and robotic setups for accurate positioning.
Scope of accreditation
Every accredited parameter, in our laboratory and at your site. CMC is the expanded uncertainty at about 95 % confidence (k = 2).
| Instrument / quantity | Discipline | Range | CMC (±) |
|---|---|---|---|
| AC Current @ 50 Hz | Electro-TechnicalAC (Measure) | 1 A to 10 A | 0.25 % to 0.27 % |
| AC Current @ 50 Hz | Electro-TechnicalAC (Measure) | 1 mA to 100 mA | 0.46 % to 0.17 % |
| AC Current @ 50 Hz | Electro-TechnicalAC (Measure) | 100 µA to 1000 µA | 1.35 % to 0.46 % |
| AC Current @ 50 Hz | Electro-TechnicalAC (Measure) | 100 mA to 1 A | 0.17 % to 0.25 % |
| AC Voltage @ 50 Hz | Electro-TechnicalAC (Measure) | 1 mV to 100 mV | 5.8 % to 0.13 % |
| AC Voltage @ 50 Hz | Electro-TechnicalAC (Measure) | 1 V to 750 V | 0.11 % to 0.12 % |
| AC Voltage @ 50 Hz | Electro-TechnicalAC (Measure) | 100 mV to 1 V | 0.13 % to 0.11 % |
| AC High Current @ 50 Hz | Electro-TechnicalAC (Source) | 100 A to 1000 A | 1.26 % to 1.33 % |
| DC Current | Electro-TechnicalDC (Measure) | 1 A to 10 A | 0.2 % to 0.19 % |
| DC Current | Electro-TechnicalDC (Measure) | 1 mA to 100 mA | 0.8 % to 0.08 % |
| DC Current | Electro-TechnicalDC (Measure) | 10 µA to 1000 µA | 3 % to 0.8 % |
| DC Current | Electro-TechnicalDC (Measure) | 100 mA to 1 A | 0.08 % to 0.2 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Measure) | 1 kΩ to 1000 kΩ | 0.01 % to 0.08 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Measure) | 1 MΩ to 100 MΩ | 0.08 % to 0.94 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Measure) | 1 Ω to 1000 Ω | 0.5 % to 0.01 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Measure) | 100 MΩ to 1 GΩ | 0.94 % to 2.3 % |
| DC Voltage | Electro-TechnicalDC (Measure) | 1 mV to 100 mV | 0.75 % to 0.3 % |
| DC Voltage | Electro-TechnicalDC (Measure) | 1 V to 1000 V | 1.5 % to 0.04 % |
| DC Voltage | Electro-TechnicalDC (Measure) | 100 mV to 1 V | 0.3 % to 1.5 % |
| DC High Current | Electro-TechnicalDC (Source) | 100 A to 1000 A | 1 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 1 kΩ to 10 kΩ | 1.2 % to 1.54 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 1 Ω to 1 kΩ | 1.85 % to 1.2 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 10 kΩ to 100 kΩ | 1.54 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 100 kΩ to 1000 kΩ | 1.54 % to 2.9 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 2 MΩ | 4.1 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 20 MΩ | 2.4 % |
| DC Resistance - 2 Wire | Electro-TechnicalDC (Source) | 200 MΩ | 3 % |
| DC Resistance - 4 Wire | Electro-TechnicalDC (Source) | 1 mΩ to 100 mΩ | 8.5 % to 2.2 % |
| DC Resistance - 4 Wire | Electro-TechnicalDC (Source) | 100 mΩ to 1 Ω | 2.2 % to 1.85 % |
| RTD (PT-100) | Electro-TechnicalTemperature Simulation (Source) | -200 °C to 600 °C | 0.87 °C |
| Thermocouple J Type | Electro-TechnicalTemperature Simulation (Source) | 50 °C to 750 °C | 0.92 °C |
| Thermocouple K Type | Electro-TechnicalTemperature Simulation (Source) | 50 °C to 1300 °C | 1.24 °C |
| Thermocouple R Type | Electro-TechnicalTemperature Simulation (Source) | 200 °C to 1700 °C | 1.62 °C |
| Thermocouple S Type | Electro-TechnicalTemperature Simulation (Source) | 200 °C to 1700 °C | 1.96 °C |
| Time | Electro-TechnicalTime & Frequency (Measure) | 1 s to 3600 s | 5.33 % to 2.44 % |
| Frequency | Electro-TechnicalTime & Frequency (Source) | 45 Hz to 999.9 Hz | 0.58 % to 0.3 % |
| Centrifuge - Non Contact Type | MechanicalAcceleration & Speed | 100 rpm to 2000 rpm | 1.8 % |
| Centrifuge - Non Contact Type | MechanicalAcceleration & Speed | 2000 rpm to 38000 rpm | 0.5 % |
| Tachometer, RPM Meter with Sensor - Contact Type | MechanicalAcceleration & Speed | 100 rpm to 3000 rpm | 1.8 % |
| Tachometer, RPM Meter with Sensor - Non Contact Type | MechanicalAcceleration & Speed | 100 rpm to 2000 rpm | 1.8 % |
Ask us. We may still be able to help.