Multimeters & Clamp Meters
Digital multimeters and clamp meters: AC/DC voltage, current (up to 1000 A) and resistance.

Multimeters and clamp meters are checked on every function and range against a multifunction calibrator: AC and DC voltage, AC and DC current and resistance. High currents up to 1000 A are generated with a current coil for clamp meters.
We can calibrate in our lab or at your site.
What we calibrate
- Digital multimeters
- Clamp meters up to 1000 A
- AC / DC voltage and current
- Resistance
- Lab or on-site
- Calibration certificate
Scope of accreditation
Ranges and capability
From our NABL scope. CMC is the expanded uncertainty at about 95 % confidence (k = 2).
| Instrument / quantity | Where | Range | CMC (±) |
|---|---|---|---|
| AC Current @ 50 Hz | In our lab | 1 A to 10 A | 0.25 % to 0.27 % |
| AC Current @ 50 Hz | In our lab | 1 mA to 100 mA | 0.46 % to 0.17 % |
| AC Current @ 50 Hz | In our lab | 100 µA to 1000 µA | 1.35 % to 0.46 % |
| AC Current @ 50 Hz | In our lab | 100 mA to 1 A | 0.17 % to 0.25 % |
| AC Voltage @ 50 Hz | In our lab | 1 mV to 100 mV | 5.8 % to 0.13 % |
| AC Voltage @ 50 Hz | In our lab | 1 V to 750 V | 0.11 % to 0.12 % |
| AC Voltage @ 50 Hz | In our lab | 100 mV to 1 V | 0.13 % to 0.11 % |
| AC High Current @ 50 Hz | In our lab | 100 A to 1000 A | 1.26 % to 1.33 % |
| DC Current | In our lab | 1 A to 10 A | 0.2 % to 0.19 % |
| DC Current | In our lab | 1 mA to 100 mA | 0.8 % to 0.08 % |
| DC Current | In our lab | 10 µA to 1000 µA | 3 % to 0.8 % |
| DC Current | In our lab | 100 mA to 1 A | 0.08 % to 0.2 % |
| DC Voltage | In our lab | 1 mV to 100 mV | 0.75 % to 0.3 % |
| DC Voltage | In our lab | 1 V to 1000 V | 1.5 % to 0.04 % |
| DC Voltage | In our lab | 100 mV to 1 V | 0.3 % to 1.5 % |
| DC High Current | In our lab | 100 A to 1000 A | 1 % |
| AC Current @ 50 Hz | On-site | 1 A to 10 A | 0.25 % to 0.27 % |
| AC Current @ 50 Hz | On-site | 1 mA to 100 mA | 0.46 % to 0.17 % |
| AC Current @ 50 Hz | On-site | 100 µA to 1000 µA | 1.35 % to 0.46 % |
| AC Current @ 50 Hz | On-site | 100 mA to 1 A | 0.17 % to 0.25 % |
| AC Voltage @ 50 Hz | On-site | 1 mV to 100 mV | 5.8 % to 0.13 % |
| AC Voltage @ 50 Hz | On-site | 1 V to 750 V | 0.11 % to 0.12 % |
| AC Voltage @ 50 Hz | On-site | 100 mV to 1 V | 0.13 % to 0.11 % |
| DC Current | On-site | 1 A to 10 A | 0.2 % to 0.19 % |
| DC Current | On-site | 1 mA to 100 mA | 0.8 % to 0.08 % |
| DC Current | On-site | 10 µA to 1000 µA | 3 % to 0.8 % |
| DC Current | On-site | 100 mA to 1 A | 0.08 % to 0.2 % |
| DC Voltage | On-site | 1 mV to 100 mV | 0.75 % to 0.3 % |
| DC Voltage | On-site | 1 V to 1000 V | 1.5 % to 0.04 % |
| DC Voltage | On-site | 100 mV to 1 V | 0.3 % to 1.5 % |
Need your multimeters & clamp meters calibrated?